X-ray Micro CT Scanner
(Model: Harmony 90H-P3-5)
A. Configurations
1. Shield Cabinet : 1 set |
2. X-ray Source : 1 set |
3. Detector : 1 set |
4. Workstation : 1 set |
5. Software : 1 set |
6. Documents : 1 set |
7. Accessories : 1 set |
B. Specifications
1. System
- Dimension: Approx. 860 x 520 x 420 (D x W x H) mm3
- Weight: approx. 140kg
- Radiation Shield Cabinet including Lead Glass Window: ¡Â 1uSv/h @ Full kVp/mA
2. Field of Views (=FOV)
- FOV variable by magnification
- Max. Magnification by geometry: 30
- Maximum FOV : Dia. 80mm.
- Stroke: Max. 239mm
- Distance of Source to Detector : 300mm
3. X-ray Source
- Integrated X-ray tube and Power supply
- Tube Voltage Range: 20~90kV
- Anode Current: 0~200uA
- Max. Power: 8watts
- Focal Spot Size: Min.5um (nominal per IEC336, JISZ47045, NEMA XR5)
- Target Material: Tungsten
- Control through RS232C
- Operating Temperature / Humidity: 10~35¡É / Max.95%
4. Flat Panel X-ray Detector
- Detector element: 1024 x 2048 pixels
- Pixel Pitch : 48um
- Active Area; 49.2 x 98.3 mm
- 12 bit data output
- Interface Cable length: Standard 3m
5. Recon. Image size
- 512 x 512 x 512, 1024 x 1024 x 1024
- Reconstruction time@ 512 x 512 x 512 , 400views < 10minnutes
- Spatial Resolution > 5um
6. Motion Axis
- Number of Motion: 2 axis(Rotation, Holizontal linear motion)
- Rotation for an object to be scanned
- Maximum Object Weight load on Rotation Stage: 3kgf
- View Number: 400, 800 selectable
- Horizontal movement for magnification: Stroke Max.239mm
7. Workstation
- 2 CPU, 64 bit Quad-Core Intel(R) Xeon(R) E5345 2.33GHz/1333Mhz/2 x 4M L2 Cache
- OS: MS windows Vista 64bit
- RAM: 16GB(2X8GB) Quad-channel DDR2 full buffer DIMM (FBD)667MHz ECC
- Hard Disk: 500GB SATA (7,200RPM) x 2EA, RAID 1.
- Optical Device: 16X max DVD+/-RW Drive
- Vidio Card: 256MB PCI , Dual_DVI ports.
- Monitor: 20" UltraSharp¢â Digital LCD Monitor 1600x1200, 1 EA
- Internal Speakers, Mouse, Key board
- Frame Grabber for x-ray detector
8. Software
- CT Software (Harmony90H): 1 set, including
1) Application program for system operating
2) For calibration
3) For CT image reconstruction
- 3D Software (iCAT3D): 1 set, including
1) Rendering: MRP/MIP/minIP/VR
2) View: Pan/Rotate/Zoom/Windowing/Invert etc
3) Measurement
Ruler/ Tape line(Line type, Curve type) /
Angle(3-Point Click, 4-Point Click, 3D Angle)
Profile: graph of intensity
Area(Line pen, Curve pen, Smart pen), Information of ROI
4) Segmentation
5) Fine Tuning
6) Output file: bmp, jpg, tif, png/avi/DICOM
9. Documents supplied
- Instruction / Operation Manual: 1set
10. Accessories
- Phantom for Rotation Center Calibration: 1 set
C. Utilities : User¡¯s preparations
- Required Electric power: single phase 220VAC(within 10%), 2kVA
- Grounding: D class grounding (Less than 100 Ohms)
- Environmental condition (in operation)
Temperature: 10 to 30 ¡É
Humidity: Less than 80% (no dew condensation)
Vibration, dust and gas to be free